Screw gauge experiment: the diameter of a wire
On this page
The standard screw gauge practical measures the diameter of a thin wire. This guide follows it step by step, with a complete observation table taken on the simulator's screw gauge, so every number in it is real arithmetic you can check, then shows the result with its uncertainty and the thickness of a sheet.
Aim and apparatus
Aim: to measure the diameter of a given wire using a screw gauge.
Apparatus: a screw gauge, a piece of wire, a half-metre scale (for the length) and a magnifying glass to read the scales.
Theory
- Pitch = distance moved by the spindle ÷ number of complete turns of the thimble. For this gauge, 1 mm.
- Least count = pitch ÷ number of circular-scale divisions = 1 mm ÷ 100 = 0.01 mm.
- Observed reading = linear scale reading (L.S.R.) + circular scale reading (C.S.R.). Multiply the division on the reference line (n) by the least count to get C.S.R. in millimetres.
- Corrected reading = observed reading − zero error. Keep the sign of the zero error. If your table lists the zero correction instead, add it to the observed reading.
The reading guide and zero error guide explain each step with figures.
Procedure
- Find the pitch and the least count of the gauge.
- Clean the measuring faces. With the gap empty, close gently with the ratchet and note the zero error with its sign. Reopen and repeat for three checks in this example, then take their mean.
- Open the gauge and place the wire across the anvil. Move the spindle close with the thimble, then finish closing with the ratchet until it clicks.
- Note the linear scale reading and the circular division on the reference line.
- Open the gauge, rotate the wire a quarter turn about its length at the same place, then close with the ratchet and take the second reading.
- Repeat both directions at each place along the wire. This example uses 5 places. Follow your lab sheet for the number it asks for.
- Work out each observed reading. Subtract the mean zero error, keeping its sign, then average the corrected readings.
Screw gauge observation table
Pitch = 1 mm; number of circular-scale divisions = 100; least count = 0.01 mm.
Zero error
| Check | Circular division on the line | Zero error |
|---|---|---|
| Zero check 1 | 2 | +0.02 mm |
| Zero check 2 | 2 | +0.02 mm |
| Zero check 3 | 2 | +0.02 mm |
Mean zero error e = +0.02 mm; zero correction = −e = −0.02 mm.
Diameter of the wire
| Reading and place | Main scale reading | Circular scale division (n) | n × least count | Observed reading | Zero error | Corrected reading |
|---|---|---|---|---|---|---|
| 1. Place 1, first direction | 0 | 73 | 0.73 | 0.73 | +0.02 | 0.71 |
| 2. Place 1, turned 90° | 0 | 74 | 0.74 | 0.74 | +0.02 | 0.72 |
| 3. Place 2, first direction | 0 | 73 | 0.73 | 0.73 | +0.02 | 0.71 |
| 4. Place 2, turned 90° | 0 | 74 | 0.74 | 0.74 | +0.02 | 0.72 |
| 5. Place 3, first direction | 0 | 73 | 0.73 | 0.73 | +0.02 | 0.71 |
| 6. Place 3, turned 90° | 0 | 74 | 0.74 | 0.74 | +0.02 | 0.72 |
| 7. Place 4, first direction | 0 | 73 | 0.73 | 0.73 | +0.02 | 0.71 |
| 8. Place 4, turned 90° | 0 | 73 | 0.73 | 0.73 | +0.02 | 0.71 |
| 9. Place 5, first direction | 0 | 73 | 0.73 | 0.73 | +0.02 | 0.71 |
| 10. Place 5, turned 90° | 0 | 74 | 0.74 | 0.74 | +0.02 | 0.72 |
The circular-scale division n is a count. Multiply it by 0.01 mm per division to get the circular scale reading. The scale readings, zero error and corrected diameter are in millimetres.
Calculation and result
The mean of the 10 corrected readings, rounded to the scale step, is 0.71 mm. The range is 0.01 mm. Half the range is 0.005 mm, so the least count of 0.01 mm sets the uncertainty for this example. The simulator uses a classroom estimate: take the larger of one least count and half the range, then round up to the scale's decimal places. Here that is ±0.01 mm, about 1.4 % of the mean.
D = 0.71 ± 0.01 mm
To estimate volume, treat the wire as a cylinder with a circular cross-section and roughly uniform diameter. Measure its length l with a ruler and use V = πD²l ÷ 4. For a length of 100 mm with an assumed uncertainty of ±1 mm, V = π × 0.71² × 100 ÷ 4, about 39.6 mm³ before final rounding.
To estimate how far the result could vary for small changes in diameter and length, add their percentage uncertainties, counting the diameter's percentage twice because D is squared. Here, 2 × 1.4 % + 1 % gives about 3.8 %, or ±1.5 mm³. Round the uncertainty to one significant figure and the volume to the same place.
V = 40 ± 2 mm³
Thickness of a sheet
To measure thickness, place the sheet's flat sides against the two measuring faces. Measure at several places. With a separate zero check showing no error, the simulator's sheet gives 2.03, 2.03, 2.04, 2.02, 2.04 mm. The mean thickness is 2.03 mm with an estimated uncertainty of ±0.01 mm. For a thin plate with an irregular outline, called an irregular lamina, use graph paper to estimate its area. If its thickness is roughly uniform, multiply the area by the mean thickness to estimate volume.
Precautions and sources of error
- Use the ratchet for the final contact and stop when it clicks. This limits the force on the wire.
- Approach every reading by closing the gauge. If you reverse at contact, close again to take up slack and avoid backlash error.
- Record the sign of the zero error as well as its size.
- Measure in perpendicular directions at several places along the wire.
- Keep your eye square to the reference line (parallax), and keep the faces clean.
Sources of error include friction in the screw, slack (backlash) in a worn screw, circular-scale divisions that are not all equal, and a wire that is not uniform.
Do it in the simulator
Open the empty screw gauge with a +0.02 mm zero error.
- Hold Turn ratchet until the faces meet. Select Record reading in the notebook to log the zero check. Open and close again for each repeat.
- Choose Copper wire, then select Close with the ratchet. Read the scales and select Record reading.
- Select Turn it 90°, close with the ratchet again and record the next reading. Then select Next place and repeat both directions. This gives 2 readings at each of the 5 places.
The observation table shows the corrected readings, their mean and an uncertainty estimate. Select Copy table to paste the data into a spreadsheet.
Questions and viva
What is the aim of the screw gauge experiment?
To measure the diameter of a given wire with a screw gauge (and from it, with the length measured by a scale, the volume of the wire). Related practicals measure the thickness of a sheet and the volume of an irregular lamina.
Why is the wire measured at different places and in two directions?
A real wire is not perfectly round or even. Readings in two directions at right angles at several places along it show how much it varies, and their mean is a better value of the diameter than any one reading.
Why is the ratchet used instead of the thimble?
The ratchet slips with a click at contact. It limits the measuring force and helps repeat it from one reading to the next. Too much force can flatten a soft wire and give a reading that is too small.
How is the zero error used in the observation table?
Find the zero error three times and take the mean (here +0.02 mm). The zero correction is minus that (−0.02 mm); add it to every observed reading.
What are the precautions in the screw gauge experiment?
Clean the faces and record the zero error with its sign. Approach each reading by closing with the ratchet. If you turn back at contact, close gently again to take up any slack. Look straight at the scales and repeat the reading in perpendicular directions at several places.
Sources
- Class XI Physics Laboratory Manual, Experiment 2, National Council of Educational Research and Training (mirrored PDF). Checked 22 November 2026.
- To measure diameter of a given wire using screw gauge, LearnCBSE.in. Checked 22 November 2026.
This site is independent and is not affiliated with any maker of measuring instruments.