Screw Gauge Simulator

Zero error in a screw gauge

On this page
  1. How to check for zero error
  2. Positive zero error
  3. Negative zero error
  4. Correcting a reading
  5. Solved examples
  6. Sign mistakes to avoid
  7. Practise zero error
  8. Questions about zero error
  9. Sources

Close the faces of a perfect screw gauge and its circular-scale zero lies exactly on the reference line. When it does not, the gauge has a zero error and every reading is off by the same amount. This guide shows how to check for it, how to read a positive and a negative zero error, and how to correct a reading.

How to check for zero error

  1. Wipe the faces of the anvil and spindle clean. A speck of dust between them shows up as a zero error.
  2. Close the faces with the ratchet until it clicks, never with the thimble alone.
  3. Look straight at the reference line. If the circular-scale zero is on it, and the edge is exactly at the main-scale zero, there is no zero error.
  4. If not, read which circular-scale line is on the reference line, and work out the error as below. Record it three times and take the mean.

Positive zero error

As the faces close, the circular-scale numbers fall towards zero. If the faces meet before the zero reaches the reference line, the zero stops short of it (below the line on these drawings), a small number lies on the line, and the main-scale zero still shows. The gauge reads more than zero, by that number of divisions:

positive zero error = + (line on the reference line × least count)

A screw gauge closed with a positive zero error: line 3 on the reference line, +0.03 mm.
School gauge: line 3 is on the reference line, so the zero error is +3 × 0.01 = +0.03 mm.
A micrometer closed with a positive zero error: line 3 on the reference line, +0.03 mm.
Micrometer: line 3 on the line, so +0.03 mm.

Negative zero error

If the zero goes past the reference line before the faces meet (above the line on these drawings), a number near the end of the scale lies on the line, and the thimble edge covers the main-scale zero. The gauge reads less than zero, so you count back from a whole turn:

negative zero error = − (number of divisions − line on the reference line) × least count

A screw gauge closed with a negative zero error: line 97 on the reference line, −0.03 mm.
School gauge: line 97 is on the line, so the zero error is −(100 − 97) × 0.01 = −0.03 mm.
A micrometer closed with a negative zero error: line 47 on the reference line, −0.03 mm.
Micrometer: line 47 on the line, so −(50 − 47) × 0.01 = −0.03 mm.

Go by the reading, not by above or below. A gauge whose thimble is numbered the other way round stops its zero on the other side of the line, so NCERT's lab manual defines the sign by what the scales read with the faces together: more than zero is a positive zero error, less than zero a negative one.

Correcting a reading

corrected reading = observed reading − zero error

Keep the sign. A positive zero error is subtracted; a negative zero error is subtracted too, which adds its size. Some books use the zero correction, the zero error with its sign changed, and add it: corrected reading = observed + zero correction. Both give the same answer.

A screw gauge with a −0.03 mm zero error reading 8.43 mm on an object.
With a −0.03 mm zero error the scales read 8.43 mm; the corrected reading is 8.46 mm.

Solved examples

  1. A screw gauge has a zero error of +0.04 mm. On a wire, the main scale reads 1 mm and circular division 50 is on the reference line. What is the corrected reading?

    Observed reading = 1 + (50 × 0.01) = 1.50 mm. Corrected reading = 1.50 − (+0.04) = 1.46 mm.

  2. A screw gauge has a zero error of −0.04 mm. On a wire, the main scale reads 1 mm and circular division 42 is on the reference line. What is the corrected reading?

    Observed reading = 1 + (42 × 0.01) = 1.42 mm. Corrected reading = 1.42 − (−0.04) = 1.46 mm.

  3. A micrometer has a zero error of +0.02 mm. On a wire, the main scale reads 7 mm and circular division 37 is on the reference line. What is the corrected reading?

    Observed reading = 7 + (37 × 0.01) = 7.37 mm. Corrected reading = 7.37 − (+0.02) = 7.35 mm.

  4. A screw gauge has a zero error of −0.03 mm. On a wire, the main scale reads 8 mm and circular division 43 is on the reference line. What is the corrected reading?

    Observed reading = 8 + (43 × 0.01) = 8.43 mm. Corrected reading = 8.43 − (−0.03) = 8.46 mm.

Sign mistakes to avoid

  • Reading a negative zero error as positive. Line 97 does not mean +0.97 mm: the zero has gone past the line and the thimble edge covers the main-scale zero, so it is −0.03 mm.
  • Adding a positive zero error. A gauge that reads +0.03 mm with its faces closed reads too much; take the error away.
  • Forgetting the error altogether. Check zero before every practical, not once a term.

Practise zero error

Questions about zero error

What is zero error in a screw gauge?

When the clean faces are closed with the ratchet, the zero of the circular scale should lie on the reference line, with the main-scale zero just showing. If it does not, the gauge has a zero error, and every reading is off by that amount.

How do you know if a zero error is positive or negative?

Close the clean faces with the ratchet and read the scales. If the gauge reads more than zero, the error is positive: the circular-scale zero stopped short of the reference line, the main-scale zero still shows, and a small number is on the line (line 3 means +0.03 mm). If it reads less than zero, the error is negative: the zero went past the line, the thimble edge covers the main-scale zero, and a number near 100 is on the line (line 97 means −(100 − 97) × 0.01 = −0.03 mm).

What is the formula for the corrected reading?

Corrected reading = observed reading − zero error, keeping the sign of the zero error. Zero correction is the zero error with its sign changed, and is added: corrected reading = observed reading + zero correction.

Why take the zero error three times?

Dirt, the force used and the way the faces meet can change a zero reading slightly. A school practical records it three times and uses the mean.

Sources

  1. Class XI Physics Laboratory Manual, Experiment 2, National Council of Educational Research and Training (mirrored PDF). Checked 22 November 2026.
  2. To measure diameter of a given wire using screw gauge, LearnCBSE.in. Checked 22 November 2026.

This site is independent and is not affiliated with any maker of measuring instruments.

Next stepThe experimentThe screw gauge practical step by step: aim, apparatus, pitch and least count, zero error, readings at five places, observation table and the result.

All guides